Abstract

Gamma-ray spectral analyses with a 5-cm × 5-cm LaBr3(Ce) detector and a NaI(Tl) detector of the same size show that the LaBr3(Ce) has much better gamma-ray peak resolution and full-energy peak counting efficiency but worse detection sensitivity. The LaBr3(Ce) detector has relatively high intrinsic radiation background due to the naturally occurring La radioisotope in lanthanum. Although this La background is entirely below the energy of 1,500 keV, additional background is in the energy region between 1,500 keV and 2,750 keV. The manufacturer attributes this radiation to alpha particles emitted by the five short-lived progeny of an Ac impurity. Comparative values for peak resolution, full-energy peak counting efficiency, and detection sensitivity are reported for Am, Co, and Cs. Results of counting Cs sources at two activity levels demonstrate the impact of background on detection sensitivity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.