Abstract

A series of tests were performed at Oak Ridge National Laboratory (ORNL) to evaluate and characterize the radiological response of a ''Cricket'' radiation detection system. The ''Cricket'' is manufactured by RAD/COMM Systems Corp., which is located in Ontario, Canada. The system is designed to detect radioactive material that may be contained in scrap metal. The Cricket's detection unit is mounted to the base of a grappler and monitors material, while the grappler's tines hold the material. It can also be used to scan material in an attempt to isolate radioactive material if an alarm occurs. Testing was performed at the Environmental Effects Laboratory located at ORNL and operated by the Engineering Science and Technology Division. Tests performed included the following: (1) Background stability, (2) Energy response using {sup 241}Am, {sup 137}Cs, and {sup 60}Co, (3) Surface uniformity, (4) Angular dependence, (5) Alarm actuation, (6) Alarm threshold vs. background, (7) Shielding, (8) Response to {sup 235}U, (9) Response to neutrons using unmoderated {sup 252}Cf, and (10) Response to transient radiation. This report presents a summary of the test results. Background measurements were obtained prior to the performance of each individual test.