Abstract

Oak Ridge National Laboratory conducted field radiological measurements at two port locations at the request of the Environmental Protection Agency (EPA). The radiological measurements were performed on five radiation detection systems at the port of Darrow, Louisiana and three systems at the port of Charleston, South Carolina. Darrow was visited on January 20-23, 2004 and Charleston on May 25, 2004. All tested systems are designed to detect radioactive material that might be present in scrap metals as the scrap is being unloaded from ships. All eight systems are commercially known as the Cricket and manufactured by RAD/COMM Systems. Each radiation detection system consists of a detector with two channels and a wireless transmitter, both mounted on the grapple, and a controller located in the crane cab. The cranes at both locations are operated by the Cooper T. Smith Company. The purpose of the radiological measurements was to evaluate the performance of the radiation detection systems in terms of their ability to detect elevated radiation levels, and to develop a routine testing method for all EPA Cricket systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.