Abstract
Magnetic resonance imaging of short relaxation time spin systems has been a widely discussed topic with serious clinical applications and led to the emergence of fast imaging ultra-short echo-time sequences. Nevertheless, these sequences suffer from image blurring, due to the related sampling point spread function and are highly prone to imaging artefacts arising from, e.g., chemical shifts or magnetic susceptibilities. In this paper, we present a concept of spherical quasi-random single-point imaging. The approach is highly accelerateable, due to intrinsic undersampling properties and capable of strong metal artefact suppression. Imaging acceleration is achieved by sampling of quasi-random points in -space, based on a low-discrepancy sequence, and a combination with non-linear optimization reconstruction techniques [compressed sensing (CS)]. The presented low-discrepancy trajectory shows ideal noise like undersampling properties for the combination with CS, leading to denoised images with excellent metal artefact reduction. Using eightfold undersampling, acquisition time of a few minutes can be achieved for volume acquisitions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.