Abstract
A new approach to quantitative analysis of major element levels by secondary ion mass spectrometry is reported. A minor isotope of the element of interest is implanted directly into the analytical sample to be utilized as a standard. This approach has been used for analysis of hydrogen in amorphous silicon. The hydrogen content determined by reference to the implanted deuterium standard was compared to a nuclear reaction analysis of the same sample; the results agreed within the combined uncertainties of the measurements (10%). Isotope fractionation in the ion ejection process was determined by analysis of a silicon sample implanted with the species HD +.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods In Physics Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.