Abstract
In the development of medical ultrasound techniques, fast and accurate pressure field measurement is important. The most common method to measure an ultrasound pressure field is mechanically scanning a hydrophone, which takes a long time and might disturb the acoustic field. In this study, we used an optical shadowgraph method. To perform this method quantitatively, it is important to define the optical propagation length precisely. For this purpose, a holographic diffuser was used as the imaging screen. Combined with a computed tomography (CT) algorithm, a pressure field was reconstructed, and the result was compared with that of hydrophone measurement. By using two shadowgraph data from short and long propagation lengths, the pressure field was successfully reconstructed even at a pressure level for high intensity focused ultrasound (HIFU) treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.