Abstract

This paper demonstrates how secondary electron energy spectroscopy (SEES) performed inside a scanning electron microscope (SEM) can be used to map sample atomic number and acquire bulk valence band density of states (DOS) information at low primary beam voltages. The technique uses an electron energy analyser attachment to detect small changes in the shape of the scattered secondary electron (SE) spectrum and extract out fine structure features from it. Close agreement between experimental and theoretical bulk valance band DOS distributions was obtained for six different test samples, where the normalised root mean square deviation ranged from 2.7 to 6.7%. High accuracy levels of this kind do not appear to have been reported before. The results presented in this paper point towards SEES becoming a quantitative material analysis companion tool for low voltage scanning electron microscopy (LVSEM) and providing new applications for Scanning Auger Microscopy (SAM) instruments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.