Abstract

A specific high-performance liquid chromatographic method has been developed for quantitative analysis of indapamide in solid pharmaceutical dosage forms on the Turkish market. A LiChrospher RP-18 column, HP-1050 UV-detector and acetonitrile-distilled water (35:65) as the solvent system were used. The flow-rate was 1.5 mL min−1. Quantitative determination of indapamide was carried out at 254 nm. The method was found to be rapid (under 10min per assay) and was selective enough to allow the determination of indapamide in the presence of certain excipients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.