Abstract

This pilot study compared the detectability of internal thermal marks produced with MRI-guided focused ultrasound (MRgFUS) on MRI, computed tomography (CT), ultrasonography (US), and color images from digital scanning. Internal marks made using MRgFUS could potentially guide surgical, biopsy or radiotherapy procedures. New Zealand White rabbits (n = 6) thigh muscle were marked using a Philips MRgFUS system. Before and after sonications, rabbits were imaged using T1- and T2-weighted MRI. Then rabbits were sacrificed and imaging was performed using CT and US. After surgical excision specimens were scanned for color conspicuity analysis. Images were read by a radiologist and quantitative analysis of signal intensity was calculated for marks and normal muscle. Of a total of 19 excised marks, approximately 79%, 63%, and 62% were visible on MRI, CT, and US, respectively. The average maximum temperature elevation in the marks during MRgFUS was 39.7 ± 10.1 °C, and average dose diameter (i.e., the diameter of the area that achieved a thermal dose greater than 240 cumulative equivalent minutes at 43 °C) of the mark at the focal plane was 7.3 ± 2.1 mm. On MRI the average normalized signal intensities were significantly higher in marks compared to normal muscle (p < 0.05). On CT, the marked regions were approximately 10 HU lower than normal muscle (p < 0.05). The results demonstrate that MRgFUS can be used to create internal marks that are visible on MRI, CT and US.