Abstract
BackgroundPhase contrast (PC) cardiovascular magnetic resonance (CMR) imaging with parallel imaging acceleration is established and validated for measuring velocity and flow. However, additional acceleration to further shorten acquisition times would be beneficial in patients with complex vasculature who need multiple PC-CMR measurements, especially pediatric patients with higher heart rates.MethodsPC-CMR images acquired with compressed sensitivity encoding (C-SENSE) factors of 3 to 6 and standard of care PC-CMR with sensitivity encoding (SENSE) factor of 2 (S2) acquired as part of clinical CMR examinations performed between November 2020 and January 2021 were analyzed retrospectively. The velocity and flow through the ascending aorta (AAo), descending aorta (DAo), and superior vena cava (SVC) in a transverse plane at the level of pulmonary artery bifurcation were compared. Additionally, frequency power distribution and dynamic time warp distance were calculated for these acquisitions. To further validate the adequate temporal resolution requirement, patients with S2 PC-CMR in the same acquisition plane were added in frequency power distribution analysis.ResultsTwenty-eight patients (25 males; 15.9 ± 1.9 years; body surface area (BSA) 1.7 ± 0.2 m2; heart rate 81 ± 16 bpm) underwent all five PC-CMR acquisitions during the study period. An additional 22 patients (16 males; 17.5 ± 7.7 years; BSA 1.6 ± 0.5 m2; heart rate 91 ± 16 bpm) were included for frequency power spectrum analysis. As expected, scan time decreased with increasing C-SENSE acceleration factor = 3 (37.5 ± 6.5 s, 26.4 ± 7.6%), 4 (28.1 ± 4.9 s, 44.7 ± 5.6%), 5 (21.6 ± 3.6 s, 57.6 ± 4.4%), and 6 (19.1 ± 3.2 s, 62.3 ± 4.2%) relative to SENSE = 2 (51.3 ± 10.1 s) PC-CMR acquisition. Mean peak velocity, net flow, and cardiac output were comparable (p > 0.87) between the five PC-CMR acquisitions with mean differences less than < 4%, < 2%, and < 3% respectively. All individual blood vessels showed a non-significant dependence of difference in fmax99 (< 4 Hz, p > 0.2), and dynamic time warp distance (p > 0.3) on the C-SENSE acceleration factor used. There was a strongly correlated (r = 0.74) increase in fmax99 (10.5 ± 2.2, range: 7.1–16.4 Hz) with increasing heart rate. The computed minimum required cardiac phase number was 15 ± 2.0 (range: 11–20) over the heart rate of 86 ± 15 bpm (range: 58–113 bpm).ConclusionsStroke volume, cardiac output, and mean peak velocity measurements using PC-CMR with C-SENSE of up to 6 agree with measurements by standard of care PC-CMR with SENSE = 2 and resulted in up to a 65% reduction in acquisition time. Adequate temporal sampling can be ensured by acquiring 20 cardiac phases throughout the entire cardiac cycle over a wide range of pediatric and young adult heart rates.
Highlights
Measuring cardiovascular flow is important, especially in pediatric population, for both congenital and acquired heart disease [1]
Patients We identified all pectus excavatum patients who had undergone clinically indicated cardiovascular magnetic resonance (CMR) examinations that included Phase contrast cardiovascular magnetic resonance (PC-CMR) with sensitivity encoding (SENSE) and with compressed sensitivity encoding (C-SENSE) sequences between November 2020 and January 2021
To further validate the correlation between heart rate and maximum frequency content of aortic blood flow waveform with uniform sampling over a typical range of heart rate observed in pediatric and young adult patients, standard of care PCCMR images of additional clinical patients obtained in the same acquisition plane were added in the frequency power distribution analysis component of the study
Summary
Measuring cardiovascular flow is important, especially in pediatric population, for both congenital and acquired heart disease [1]. Time-resolved phase contrast (PC) cardiovascular magnetic resonance (CMR) imaging allows measurement of blood velocities, volume flow rate, total flow, and pressure gradients in the major cardiothoracic vasculature and heart. It is considered the clinical reference standard for quantification of blood flow [2,3,4,5]. A free breathing retrospectively cardiac gated k-space segmented two-dimensional PC-CMR acquisition with a parallel imaging acceleration factor of 2 may be the preferred sequence for quantitative flow assessment over the cardiac cycle in pediatric patients, many of whom find it difficult to sustain steady respiratory suspension of adequate duration. Additional acceleration to further shorten acquisition times would be beneficial in patients with complex vasculature who need multiple PC-CMR measurements, especially pediatric patients with higher heart rates
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have