Abstract
AbstractThis paper reports on our experiences of applying S# (“safety sharp”) to model and analyze the case study “hemodialysis machine.” The S# safety analysis approach focuses on the question, what happens if we place a controller with correct software into an unreliable environment. To answer that question, the S# toolchain natively supports the Deductive Cause Consequence Analysis, a fully automatic model checking‐based safety analysis technique that determines all sets of component faults with the potential of causing a system hazard. Furthermore, S# can give an approximate estimate of the hazard's probability. To demonstrate our approach, we created a model with a simplified controller of the hemodialysis machine and relevant parts of its environment and performed a safety analysis using Deductive Cause Consequence Analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.