Abstract

IntroductionStrategies for achieving high resolution varies between manufacturers. In CT, the helical mode with narrow collimation has long been considered as the gold standard for high-resolution imaging. More recently, incremental modes with small dexels and focal spot, have been developed but have not been compared with helical acquisitions under optimal conditions. The aim of this work is to compare the high-resolution acquisition strategies currently proposed by recent MSCT. MethodsThree CT systems were compared. A phantom was used to evaluate geometric accuracy, uniformity, scan slice geometry, and spatial resolution. Human dry bones were used to test different protocols on real bone architecture. A blind visual analysis was conducted by trained CT users for classifying the different acquisitions (p-values). ResultsAll systems give satisfactory results in terms of geometric accuracy and uniformity. The in-plane MTF at 5% were respectively 13.4, 15.9 and 18.1 lp/cm. Dry-bones evaluation confirms that acquisition#3 is considered as the best. ConclusionsThe incremental acquisition coupled with à small focal spot, and a high-sampling detector, overpasses the reference of low-pitch helical acquisitions for high-resolution imaging. Cortical bone, bony vessels, and tumoral matrix analysis are the very next challenges that will have to be managed to improve normal and pathologic bone imaging thanks to the availability UHR-CT systems.