Abstract
The 5HT1D agonist sumatriptan is efficacious in the treatment of migraines. MK-462 is a drug of the same class which is under development in our laboratories. Bioanalytical methods of high efficiency, specificity and sensitivity were required to support the preclinical and clinical programs. These assays were based on HPLC with tandem MS-MS detection. MK-462 and sumatriptan were extracted using an automated solid-phase extraction technique on a C2 Varian Bond-Elut cartridge. The n-diethyl analogues of MK-462 and sumatriptan were used as internal standards. The analytes were chromatographed using reversed-phase (nitrile) columns coupled via a heated nebulizer interface to an atmospheric pressure chemical ionization source. The chromatographic run times were less than 7 min. Both methods were precise, accurate and selective down to plasma concentrations of 0.5 ng/ml. The assay for MK-462 was adapted to separately monitor the unlabeled and 14C-labeled species of the drug following intravenous administration of radiolabeled material to man.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.