Abstract
Synthetic cannabinoid receptor agonists (SCRA) share minimal structural similarities to tetrahydrocannabinol or themselves. Due to their heterogeneous structures and the rapid appearance and disappearance of new SCRA on the drug scene, the quantitation of SCRA has not been attempted extensively. We present a wide series of SCRA concentrations based on a single-point calibration using peak height ratios for the extracted ion chromatogram of the protonated precursor ion against that of the internal standard. These concentrations are viewed as indicative only given the use of a single concentration "calibrator" based on the response of a deuterated analogue of a structurally related compound. What is of note, is that, despite the potential differences in potency the majority of SCRA seem to have relatively similar concentrations in postmortem cases.