Abstract
A rapid and sensitive procedure is described for the quantitation of levamisole in plasma using high-performance liquid chromatography (HPLC). The procedure involves sample preparation using a reverse-phase C18 cartridge prior to chromatography and quantitation using peak area ratios (UV absorbance detection, 225 nm) of levamisole to the internal standard, quinine. The limit of detection was 21 ng/ml and the limit of quantification was 72 ng/ml, both contained in 1 ml of plasma. The recoveries were sufficiently high (73.1%) and overall coefficient of variation of the procedure was 0.25%. This procedure has been used to determine levamisole levels in human and cattle plasma. A comparison of using two C18 columns (Nova-pak, Puresil) was also studied and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European Journal of Drug Metabolism and Pharmacokinetics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.