Abstract
This study evaluates the potential of laser induced breakdown spectroscopy (LIBS) coupled with chemometrics to develop a quantification model for rubidium (Rb) in minced beef. A LIBSCAN 150 system was used to collect LIBS spectra of minced beef samples. Beef liver was used to spike the Rb levels in minced beef. All samples were dried, powdered and pelleted using a hydraulic press. Measurements were conducted by scanning 100 different locations with an automated XYZ sample chamber. Partial least squares regression (PLSR) was used to develop the calibration model, yielding a calibration coefficient of determination (Rc2) of 0.99 and a root mean square error of calibration (RMSEC) of 0.05ppm. The model also showed good results with leave-one-out cross validation, yielding a cross-validation coefficient of determination (Rcv2) of 0.90 and a root mean square error of cross-validation (RMSECV) of 0.22ppm. The current study shows the potential of LIBS as a rapid analysis tool for the meat processing industry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.