Abstract
We report an accurate quantification of free sialic acid (SA) in human plasma using LC-MS/MS method with isotope-labeled standard calibration (ILSC) and robust derivatization. Specifically, derivatization of SA with a stable and inexpensive 3,4-diaminotoluene (DAT) provides a stable product of SA with high MS response, proving a convenient and cost-effective LC-MS/MS analysis of free SA. In addition, the use of (13)C3-SA as calibration standard ensured the accuracy for the measurement. This assay used ultra high performance liquid chromatography (UHPLC) for separation of native/labeled SA and IS from matrix interference, and employed mass spectrometry in multiple reaction monitoring (MRM) mode for sensitive and selective detection. We have achieved a lower limit of quantification (LLOQ) of 20ng/mL and a total running time of 4.2min, which is the most sensitive and quick measurement for free SA in biomatrices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.