Abstract

The availability of reliable analytical methods for measuring amine concentrations is necessary for optimum operation of aqueous amine CO2 separation systems being employed for postcombustion capture (PCC) of CO2. A GC-FID (gas chromatography with flame ionization detection) method is described for the reliable quantification of 30% (w/w) monoethanolamine (MEA) in severely degraded solvent samples. The observation of intermittent splitting of the MEA peak was a major concern with this approach. The use of a wide-bore column led to improved MEA peak resolution and peak shape. The reliability and robustness of the GC-FID method were assessed by analyzing degraded 30% (w/w) MEA solvent samples from CSIRO’s pilot plant at AGL’s Loy Yang power station in Victoria, Australia. The results were compared with those obtained by titration and total organic carbon (TOC) measurements of the same samples. The MEA concentrations obtained by the GC-FID and titration methods were statistically similar. In contrast, the MEA concentrations calculated from TOC were consistently higher than those obtained by both GC-FID and titration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.