Abstract

A five-step model for a quality assurance system is developed for an internal quality control check. It includes the quality control of the decomposition method and the detection method as steps belonging together. The Wickbold combustion technique as decomposition method in combination with atomic absorption spectrometry was chosen. The vaporization of the elements mercury, arsenic, lead, antimony and selenium is based on combustion in an oxyhydrogen flame. To check the efficiency of the analytical system, the uncertainty of results was calculated on the basis of the "Guide to the Expression of Uncertainty in Measurement".

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.