Abstract

This document defines the procedure necessary to qualify the airborne molecular cleanliness (AMC) of vacuum systems (enclosures or large components) that are placed within the National Ignition Facility (NIF) target chamber or are attached to it and communicate with it during vacuum operation. This test is specific to the NIF target chamber because the allowable time dependent rate of rise in the pore filling of a sol-gel coated SAW sensor is based on some nominal change-out time for the disposable debris shields. These debris shields will be sol-gel coated and thus they represent a means of ''pumping'' AMCs from the target chamber. The debris shield pumping rate sets the allowable change in pore filling with time specified in the test procedure. This document describes a two-part procedure that provides both a static measurement of sol-gel pore filling at the end of a 48-hour test period and a dynamic record of pore-filling measured throughout the test period. Successful qualification of a vacuum system requires that both the static and dynamic measurements meet the criteria set forth in Section 7 of this document.