Abstract

A simple purge-and-trap gas chromatographic method with flame ionization detection was developed for the determination of styrene in urine and blood. Styrene present in a 5 ml sample at room temperature was swept by helium at 40 ml/min for 11 min, trapped on a Tenax trap, desorbed by heating, cryofocused, and injected by flash heating into a DB-5 capillary GC column. The oven temperature program was from 80°C, held for 8 min, to 120°C at 5°C/min, and then held for 2 min. The detector temperature was 250°C. The calibration curves were linear in the range of 2.5–15 ppb styrene in urine and 25–150 ppb in blood. The detection limits calculated were 0.4 μg/l in urine and 0.6 μg/l in blood. The coefficients of variations within the day and day-to-day were 3 and 3.1%, respectively, for 2.5 ppb of styrene in urine, and 1 and 1.6% for 25 ppb of styrene in blood. The results obtained from samples taken from workers exposed to styrene were reported.