Abstract
The applicability of pulsed splitless injection to the gas chromatographic analysis of pesticide residues in fruit and vegetables has been evaluated. 22 pesticides belonging to different chemical classes, including those known to be liable to matrix induced response enhancement, were selected for the study. The parameters of pressure pulse have been tested for optimum performance of injection. Application of the pressure pulse was found to decrease matrix effects during analyses of real samples. Further decline of matrix effects was obtained using higher sample injection volumes. The installation of a deactivated retention gap was necessary to obtain good peak shapes with injection volumes exceeding 1 μL of sample. Up to 4 μL was then injected without peak distortion and consequent loss of resolution. Using 4 μL pulsed splitless injection, matrix effects were almost completely eliminated even at very low concentration levels of analytes. The highest matrix effects observed for tested compounds at the lowest concentration level tested were in the range of 110–122%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.