Abstract

A pulsed eddy-current system has been developed for production inspection of small, thin-wall, non-ferromagnetic tubing. To detect and accurately size both outer and bore-side flaws required an operating frequency higher than available from present commercial equipment. A pulsed eddy-current instrument was designed and constructed that used 3.2 MHz square waves, with a bandwidth of 20 MHz. The system is able to reliably detect flaws as small as 0.015 mm (0.6 mils) on either the inner or outer surface of the tube. A computer controls the scanning of the tube, as recording, analyzing, and plotting the data. The computer programs and instrument details are given in the report.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.