Abstract

We propose in this paper novel approach for improving mass resolution in atom probe tomography. Using conventional counter-electrode or microelectrode design, improved mass resolution at half, tenth and 1% of the mass peak maximum is predicted when shaping properly the voltage evaporation pulse. Using a numerical approach, it is shown that a flat top voltage pulse used to trigger the field evaporation with sharp front and leading edges associated with a short tip to counter electrode distance(≈10μm) strongly minimizes the energy deficits of evaporated ions from the sample, so that energy compensation devices are not necessary to obtain high mass resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.