Abstract

Abstract The pulsatile flow fields downstream of the three aortic bileaflet mechanical valves: St. Jude Medical (SJM), CarboMedics (CM), Edwards Tekna (modified Duromedics, DM) have been investigated in vitro. An experimental circulatory mock loop has been used to simulate the left heart under the conditions of geometric and dynamic similarities. These tested valves have been mounted on the base of a glass‐made model aorta with symmetrical sinus. The blood analog fluid has a viscosity of 3.9 cP and has matching refractive index with the model aorta. Experimental parameters have been set at the rate of 70 heartbeats and 5 liters of flow per minute. A two‐component laser Doppler anemometer has been used to measure the distributions of pulsatile velocities at five different axial locations inside the sinus of the aorta. Results show that although all the three valves are aortic bileaflet valves, they have different flow patterns, due to their different opening angles, shapes of leaflets and movements of hing...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.