Abstract

There have been limited publications that report long-term outcomes of pyrocarbon implants. This report describes both clinical and radiographic long-term results for patients who have been treated with pyrocarbon proximal interphalangeal implants. Thirteen implants in ten patients are reported for an average follow-up of 8.3 years (range 6.2-9.3). All patients were suffering from degenerative joint disease. Five of the 13 digits were free of pain, the remaining eight digits had mild to moderate pain (visual analogue scale 2-5). The average active range of motion was 58° (SD 19°) at latest examination. X-ray results were unremarkable in six digits with an acceptable position of the prosthesis. However, in seven patients significant radiolucent lines (≥ 1 mm) were observed. Three prostheses demonstrated a migration of the proximal component, and one a subsidence of the distal component. Our study does not support the use of this implant for treatment of osteoarthritis of the finger joint owing to high complication rates and limited range of motion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.