Abstract

Introduction The aim of this article is to introduce a novel protocol for proton pencil beam scanning treatment with moderate deep inspiration breath hold (mDIBH) and report on our clinical implementation results. Methods Three computed tomography (CT) scannings to build the patient's anatomy model were performed during the patient's voluntary mDIBH. All 3 CT scans were used in the optimization during the treatment planning process. Both orthogonal kV imaging and cone-beam computed tomography (CBCT) were implemented for patient alignment with BH prior to the treatment. The BH CBCT images were analyzed for BH reproducibility and the virtual total dose (VTD) retrospectively. To find the VTD, a series of deformable image registrations (DIR) were performed between CBCT and pCT. The effect of the variation of lung density on the dose distribution was also analyzed in the study. Results The values of the mean, standard deviation, maximum, and minimum of the tumor location difference between the CBCT and pCT were 1.9, 1.6, 4.7, and 0.0 mm, respectively. The percentage difference in D99% of CTVs between VTD and the nominal plan was within 1.5%. Conclusions The feedback-based voluntary moderate BH proton PBS treatment was successfully performed in our clinic. This study shows that there is a potential to implement the BH treatment widely in proton centers.