Abstract

The purpose of the present study was to fabricate a volume coil for proton/deuterium magnetic resonance imaging (MRI) in rodents at 9.4 T. Two birdcage radiofrequency (RF) coils have been designed for proton/deuterium MRI: the rungs of two concentric birdcages were azimuthally interleaved with each other for better decoupling, and the two coils were tuned to 400.3 and 61.4 MHz for 1 H/2 H resonance at 9.4 T. Compared to a commercially available coil, the proposed 1 H/2 H RF coil provides reasonable transmission efficiency and imaging signal-to-noise ratio (SNR); the relationships among imaging parameters such as SNR, voxel size, and deuterium oxide concentrations have been quantitatively studied, and the linear correlation results together with the spectroscopic data in vivo indicate its feasibility in deuterium metabolic imaging (DMI) in vivo. Our study indicates that using the birdcage design for MRI signal excitation combined with surface coil array for signal reception can facilitate DMI investigations more effectively towards future pre-clinical and clinical applications. As a noninvasive method by measuring nonhydrogen nuclear deuterium signals to reflect metabolite information, DMI will feature prominently in future precision medicine through the whole process of diagnosis, treatment, and prognosis. © 2021 Bioelectromagnetics Society.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.