Abstract

A methodology based on the coupling of a headspace autosampler with a GC and a MS detector operating in SIM mode has been developed for the determination of volatile organic compounds (THMs and BTEX) in soils. The GC device used is equipped with a programmable temperature vaporizer (PTV) packed with Tenax-TA ® to introduce the samples (the injection mode used was solvent vent), and a modular accelerated column heater (MACH™) to control column temperature. The proposed measurement procedure reduces the sample pretreatment step to a minimum. Combined use of solvent vent injection mode and mass spectrometry detection allows a highly sensitive method to be proposed, with limits of detection of the order of ng/kg for all the target compounds. Furthermore, the capillary column used allows rapid separations of compounds in less than 4.60 min, affording a very short total analysis cycle time of 9 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.