Abstract

The output signal generated by detectors used in ion mobility spectrometry (IMS) is a time-dependent, small ionic current. To be able to take full advantage of information contained in this signal, adequate procedures of signal processing need to be utilized. Methods of spectrum filtration, peak separation, base-line correction as well as one- and two-dimensional integration applied in quantitative analysis are described. The effectiveness of the chosen methods was demonstrated on examples of experimental results obtained by IMS. Measurements were performed for spectra of reactant ions and sample ions generated by acetone, methyl tert-butyl ether (MTBE), dimethyl methylphosphonate (DMMP) and benzene. It was demonstrated that the proposed methods considerably improve the quality of the spectra, increase the selectivity of detection and reduce the uncertainty of quantitative measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.