Methyl tertiary-butyl ether (MTBE) is an organic compound which is used as a gasoline additive. Contamination of ground and surface water can occur due to large scale use of MTBE and its high solubility in water. According to United State Environmental Protection Agency (USEPA), MTBE is a possible human carcinogen at high doses and its detection and measurement in the water is important as concerned about human health. In this work, ion mobility spectrometry (IMS) equipped with a corona discharge ionization source was used for determination of MTBE in drinking water. Both pure and aqueous solutions of MTBE were studied and their ion mobility spectra were obtained at different temperatures. Using a calibration curve for detection of MTBE in drinking water, a detection limit (LOD) of 1 mg/L was obtained by IMS. This work proved that, IMS with corona discharge can be used for fast and direct detection of MTBE in water sample without any sample preparation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.