This paper presents the primary shock calibration system developed at Inmetro to obtain the shock sensitivity of accelerometers and acceleration measuring chains. The shock exciter employs a pneumatic driver and is based on the mechanical shock between two rigid bodies guided by air bearings. The system follows the ISO standard 16063-13. The shock acceleration is measured by laser interferometry. Details of the calibration system, including the shock machine, data acquisition and signals processing are presented and some measurement results are discussed by the authors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.