Abstract

A signal processing method for determining the dynamic behavior of accelerometers by shock excitation and laser interferometry is presented. The method allows the shock sensitivity and the magnitude and phase lag of the complex sensitivity of accelerometers to be accurately measured. After digitizing the phase‐modulated signals at the output of a homodyne or heterodyne interferometer, the time‐dependent displacement is reconstructed on the basis of the principle of coherent demodulation. Data processing developed for determining peak value and spectrum of the acceleration efficiently suppresses disturbing quantities. Computer simulations and experiments proved that the shock parameters can be measured with expanded uncertainties of less than 0.2%. method has been adopted by the ISO/TC 108/SC 3 as a primary shock calibration method to be specified in a new international standard ISO 16063‐13 “Primary shock calibration using laser interferometry”.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.