Abstract
Prolonged times to tracheal extubation are associated with adverse patient and economic outcomes. We simulated awakening patients from sevoflurane after long-duration surgery at 2% end-tidal concentration, 1.0 minimum alveolar concentration (MAC) in a 40-year-old. Our end-of-surgery target was 0.5 MAC, the Michigan Awareness Control Study's threshold for intraoperative alerts. Consider an anesthetist who uses a 1 liter/minute gas flow until surgery ends. During surgical closure, the inspired sevoflurane concentration isreduced from 2.05% to0.62% (i.e., MAC-awake). The estimated time to reach 0.5 MAC is 28 minutes. From a previous study, 28 minutes exceeded ≥95% of surgical closure times for all 244 distinct surgical procedures (N=23,343 cases). Alternatively, the anesthetist uses 8 liters/minute gas flow with the vaporizer atMAC-awake for 1.8 minutes, which reduces the end-tidal concentration to 0.5MAC. The anesthetist then increases the vaporizer to keep end-tidal 0.5 MAC until the surgery ends. An additional simulation shows that, compared with simulated end-tidal agent feedback control, this approach consumed 0.45 mL extra agent. Simulation results are the same for an 80-year-old patient. The extra 0.45 mL has a global warming potential comparable to driving 26 seconds at 40 kilometers (25 miles) per hour, comparable to route modification to avoid potential roadway hazards.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.