Abstract

The irradiation of blood before transfusion into immunosuppressed patients is an increasingly common technique used to prevent graft-versus-host disease. A technical procedure is described for the calibration of blood irradiators, including the determination of absolute dose rate and relative dose distribution over the blood volume. Results of dose rate measurements on commercially available irradiators indicate differences of +5% to -13% with manufacturer-supplied calibrations and variations in the relative dose rate over the irradiation volume from 70% to 180%. The clinical implications of these findings and the need for accurate dosimetry are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.