Abstract

This article reports the first use of the pressure reciprocity technique to calibrate a micro-electromechanical system (MEMS) microphone. This standardized primary calibration method is conventionally used to calibrate laboratory standard microphones. Results for the pressure reciprocity calibration of a MEMS microphone and two laboratory standard microphones are presented for the frequency range 100-10 000 Hz. Because the amplifier in the MEMS microphone package prevents reciprocal operation, this microphone was used only as a receiver of sound. A description of the procedure is presented along with checks of the measurement results and data regarding the uncertainties of these results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.