Abstract

Measurements have been made on the distribution of pressure and hysteresis effects in a tetrahedral anvil apparatus by simultaneously monitoring the resistances of calibration wires placed at various positions in the pyrophyllite tetrahedrons. A pressure gradient of about 1% is observed, and it is therefore suggested that pressure calibration wires may be placed on the tetrahedron face to determine the pressure throughout the tetrahedron volume. This technique will allow pressure calibration when sample porosity is not known and could possibly permit pressure calibration at elevated temperatures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.