Abstract

A new X-ray spectrometer has been installed in the BL08W beamline at SPring-8 for high-resolution Compton scattering measurements using 90–120 keV X-rays. The spectrometer consists of a Cauchois-type triple-crystal analyzer and a quasi position sensitive detector with four slits. The detector works not only as a position-sensitive detector but also as an energy sensitive detector, which allows for low background noise measurements. An overall resolution of 0.10 atomic units in momentum space has been achieved at the incident X-ray energy of 115 keV. Compton profiles of niobium have been measured along the [1 0 0] and [1 1 0] directions in order to demonstrate the performance of the spectrometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.