Abstract

A method was set up to prepare reference sources for calibration of surface contamination monitors using Nafion HP membrane to fix the radioactive nuclides on aluminium backing material. By investigating the adsorption mechanism including kinetics and pH, it has been proved the 241Am could be fixed on the membrane effectively. The surface emission rate and the activity was measured to characterize the sources. In addition, the uniformity of the sources was evaluated according to ISO 8769 and the standard deviation of the surface emission rates from each individual portion of the whole source was no greater than 5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.