Abstract
Abstract A new type of β-ray calibration source was designed for estimating tissue doses from skin contamination. Thin polyethylene ion-exchange membranes were applied for the source preparation. The method of determining tissue absorbed dose rates with an extrapolation chamber was discussed for the calibration sources prepared with 147Pm, 204TI and 90Sr-90Y. The influence of backscattered β-rays by source mount materials was estimated as a function of membrane source thickness. The membrane sources are quite suitable for the present purpose, because they have large uniform radioactive areas and can be easily handled without radioactive contamination.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.