A new aluminium (III) polymeric membrane electrode containing (E)-N′-(4-(dimetlylamino)benzylidene)-1H-benzo[d]midazobe-2-hydrazide (L) as a carrier has been prepared. The influence of membrane ingredients including ionophore, PVC, plasticizers and anionic excluders (NaTPB and KTpClPB) on the electrode response has been investigated. The best performance was obtained with a membrane composition of carrier L:DBP:PVC ratio of 1.00:65.99:33.01 (%, w/w). The sensor exhibited a Nernstian response for Al3+ ion ranging from 1.0×10−7 to 1.0×10−1M with a detection limit of 5.6×10−8M and a slope of 20.7mV/decade. It showed a relatively fast response time (10s), suitable reproducibility and stability, and good selectivity towards Al3+ ion. The prepared electrode was used as an indicator electrode in potentiometric titration of Al3+ with EDTA and in the determination of Al3+ ion in waste water and vermicelli.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.