Abstract

A new polymeric membrane electrode (PME) and a coated platinum disk electrode (CPtE) based on Schiff base complex of Mo(VI) as a suitable carrier for I − ion were described. The influence of membrane composition, pH and possible interfering anions were investigated on the response properties of the electrodes. The electrodes exhibited a Nernstian slope of 63.0 ± 0.5 (CPtE) and 60.3 ± 0.4 (PME) mV decade − 1 in I − ion over a wide concentration range from 7.9 × 10 − 7 to 1.0 × 10 − 1 M for CPtE and 9.1 × 10 − 6 to 1.0 × 10 − 1 M I − for PME. The potentiometric response of the electrodes was independent of the pH of the test solution in the pH range 2.0–8.5 with a fast response time (< 10 s). The process of transfer of iodide across the membrane interface was investigated by use of the AC impedance technique. The proposed sensors were successfully applied to direct determination of iodide in samples containing interfering anions, waste water and as indicator electrodes in precipitation titrations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.