Abstract
A liquid membrane ion-selective electrode has been prepared for the measurement of carbonate ion activities in solutions containing high levels of hydrogen carbonate and chloride. The electrode exhibits Nernstian response in the approximate range 1O -2–1O -7 M carbonate, and has a fast response and low noise level under most conditions. In the physiologically important pH region, hydroxide ion does not interfere. Detailed information is given on the composition and properties of the active membrane phase.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.