Abstract

For the calibration purpose of thoron ((220)Rn) measurements, solid sources with constant (220)Rn emanation are generally desired. In this study, Th(4+) in the thorium nitrate powder was successfully exchanged to the surfaces of sodium benzenesulphonate resins with a bead diameter of ∼0.5 mm by using the ion-exchange method. Tests of (220)Rn emanating from the ion-exchanged resin indicated that the emanation rates were nearly constant, provided both the air humidity and air temperature were controlled. As the resin can be easily prepared to be of various sizes and activities of sources, the ion-exchanged resin is considered as a promising standard source for calibrations of (220)Rn measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.