Abstract

The wall shear stress (WSS) contains useful information for the diagnosis of vascular diseases. To improve the accuracy of WSS measurements by ultrasound, some studies have attempted to correct the underestimation of the WSS by approximating a blood flow velocity profile measured by ultrasound to a velocity distribution model such as laminar flow. In this study, a blood flow velocity profile measured by ultrasound was modeled by the Womersley flow model, in order to correct underestimation of WSS in cases of measurements of vessels with complex geometries such as stenotic vessels. The simulation results showed that the rms error of the estimated WSSs was 45.8% from the stenotic vessel with a stenotic ratio of 40% when the spatial gradient was estimated by fitting a linear model to the measured flow profile. Such an error was improved from 45.8% to 25.3% by fitting a Womersley flow model to the measured blood velocity profile.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.