Abstract

Preliminary studies on effective and eye lens doses of six Radiologists, four Cardiologists have been conducted for a period of 3 months. Electronic dosemeters positioned under and over lead apron of staff were used for the dosimetry. The estimated effective dose per month to Cardiologist and Radiologist were 0.01-0.07 mSv and 0.03-0.14 mSv, respectively. The estimated eye lens doses per month to Cardiologists and Radiologists were also 0.15-0.30 mSv and 0.53-3.39 mSv, respectively. The effective doses per month to staff were below the ICRP acceptable limit of 1.67 mSv/month but the upper limit of the range of estimated eye lens dose exceeded the ICRP acceptable limit by a factor of 2. Regular use of protective goggles and consistent eye lens dose monitoring is encouraged at the hospital for dose optimization.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.