Abstract

AbstractIn order to reveal the dynamic response characteristic of thin film thermocouples (TFTCs), the nichrome/nisil (NiCr/NiSi) TFTCs are prepared onto the glass substrate. With short pulse infrared laser system, NiCr/NiSi TFTCs are dynamically calibrated. The thermoelectric electromotive force (TEF) curves of NiCr/NiSi TFTCs are recorded by the memory hicorder system, which could reflect TEF signals with resolution ratio in nanosecond and microvolt, simultaneously. With increasing laser energy from 15.49 to 29.59 mJ, TEF curves display more and more violent oscillation, even negative value. The results show that the bounce of thermal energy happens between two interfaces of TFTCs because the thermal conductivity of glass and air is significantly lower than that of NiSi/NiCr TFTCs. The bounce of thermal energy results in the obvious decrease of nNiCr and nNiSi, as well as oscillation of TEF. For laser energy in 29.59 mJ, the bounce of thermal energy in NiCr film could result in nNiCr < nNiSi. Then, TEF value appears abnormal negative value. Based on the results, the complex thermal energy transport process in TFTCs dynamic calibration is revealed, which results in the oscillation of thermal energy and TEF signal.