Abstract
An experimental intercomparison between the proton response of LiF TLD-100 and alanine detectors is reported. The investigations were performed with LiF chips and alanine pellets in a 62 MeV proton beam at the Paul Scherrer Institut in Villigen (CH). Results were compared with reference dosimetry provided by a Markus type parallel plate ionisation chamber. The response of the detectors was studied, in a phantom, at different beam penetration depths in pristine and modulated beams. For both alanine and TL detectors, within the experimental uncertainty of the measurements, no significant energy dependence in the response was observed down to the Bragg peak region. The sensitivity of alanine and LiF detectors to protons was measured in the centre of a modulated Bragg peak and no significant difference was found with respect to 60 Co. Contrary to LiF, alanine also offers a remarkable tissue equivalence which favours its choice for in-phantom dosimetry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.