Abstract

In helicopter emergency medical services, HEMS, coagulopathy presents both in trauma (e.g. consumption of coagulation factors) and non-trauma cases (e.g. anticoagulant use). Therefore, in HEMS coagulation measurements appear promising and Prothrombin Time (PT) and derived INR are attractive variables herein. We tested the feasibility of prehospital PT/INR coagulation measurements in HEMS. This study was performed at the Dutch HEMS, using a portable blood analyzer (i-Stat®1, Abbott). PT/INR measurements were performed on (hemodiluted) author's blood, and both trauma- and non-trauma HEMS patients. Device-related benefits of the i-Stat PT/INR system were portability, speed and ease of handling. Limitations included a rather narrow operational temperature range (16-30 °C). PT/INR measurements (n = 15) were performed on hemodiluted blood, and both trauma and non-trauma patients. The PT/INR results confirmed effects of hemodilution and anticoagulation, however, most measurement results were in the normal INR-range (0.9-1.2). We conclude that prehospital PT/INR measurements, although with limitations, are feasible in HEMS operations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.