Abstract

A procedure based on curve fitting techniques for the calculation of the retention data in linear programmed-temperature gas chromatography was applied in order to calculate the elution temperature of fifteen polycyclic aromatic hydrocarbons on a DB- 5 capillary column. The linear programmed-temperature retention data were calculated with a BASIC program starting from isothermal retention times and the calculated values were compared with experimental data. For five different linear temperature programmes the accuracy expressed as fractional difference was always better than 1% in spite of the simplifications introduced in the calculation methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.